Source avec lien : Journal of Occupational & Environmental Hygiene, 20(7). 10.1080/15459624.2023.2205481
L’objectif principal de cette étude était d’évaluer l’effet du débit du PAPR et du rythme de travail de l’utilisateur sur les performances du PAPR, à l’aide d’une méthode d’évaluation basée sur un mannequin.
Loose-fitting powered air-purifying respirators (PAPRs) are used in healthcare settings, although barriers to routine, everyday usage remain, including usability concerns and potential interference with work activities. Loose-fitting PAPRs are approved by the National Institute for Occupational Safety and Health (NIOSH) and must meet minimum performance requirements, including a minimum airflow requirement of 170 L/min. One course of action to address usability concerns is to allow for the use of PAPRs designed with reduced airflow rates. The primary objective of this study was to assess the effect of PAPR flow rate and user work rate on PAPR performance, using a manikin-based assessment method. PAPR performance was quantified using the « Manikin Fit Factor » (mFF), a ratio of the challenge aerosol concentration to the in-facepiece concentration. Flow rates from 50–215 L/min and low, moderate, and high work rates were tested. Two models of NIOSH Approved loose-fitting facepiece PAPRs were tested, both having an Occupational Safety and Health Administration Assigned Protection Factor (APF) or expected level of protection, of 25. A two-way analysis of variance with an effect size model was run for each PAPR model to analyze the effects of work rate and flow rate on PAPR performance. Flow rate and work rate were found to be significant variables impacting PAPR performance. At low and moderate work rates and flow rates below the NIOSH minimum of 170 L/min, mFF was greater than or equal to 250, which is 10 times the OSHA APF of 25 for loose-fitting facepiece PAPRs. At high work rates and flow rates below 170 L/min, mFF was not greater than or equal to 250. These results suggest that some loose-fitting facepiece PAPRs designed with a flow rate lower than the current NIOSH requirement of 170 L/min may provide respirator users with expected protection at low and moderate work rates. However, when used at high work rates, some loose-fitting facepiece PAPRs designed with lower flow rates may not provide the expected level of protection.